Type: | UN201 | Medicine Capacity: | Max 25ml |
Power: | 3.0W | Power By: | 2*AA 1.5V Battery |
Work Sound: | ≤ 50dB | Particle Size: | MMAD 4.0μm |
Weight: | About 94g | Working Temp: | 10 – 40℃ |
Medicine Temperature: | ≤50℃ | Product Size: | 67*42*116mm (2.64*1.65*4.57 inch) |
Mist Particle Size Distribution: | ≤ 5μm >65% | Nebulization Rate: | ≥ 0.25ml/min |
Function:Aerosol therapy of asthma,allergies and other respiratory disorders for hospital and home care use.
Principle of use:Ultrasonic nebulizer sprayed liquid medication to fog panel by compressing air,and form tiny particles,which flow into the throat through imbibing tube.
Characteristics:Quiet, Easy carry and clean, Have two modes to choose,Can be automatic turn off in 5 or 10minutes. Mesh nebulizer is an ideal product for all the people when they suffering asthma, allergies and other respiratory disorders.
1.The device recharges with a USB cord.
2.The LED light will be orange while charging and blue when fully charged.
3.Runtime on a full charge is approximately 120 minutes.
1.To clean accessories : remove the mouthpiece and any accessories from the device, wipe or soak with medical wipe.
2.To clean the nebulizer : add 6ml of clean water to the container cup and start the automatic cleaning mode. Remove any the mesh plate and remove any residue.
3. If the outside of the device needs cleaning, wipe with dry towel.
4.Return the mesh plate to the device after full cleaning and store in a dry and cool place.
5.Make sure to charge the battery AT LEAST every 2 month to keep battery life strong.
6. Clean the medicine cup immediately after use and don’t leave any solution in the machine, keep the medicine cup dry.
ISSUES & FAQS |
REASONS AND TROUBLESHOOTING |
There is little or no aerosol coming out of the nebulizer. | 1 Insufficient liquid in the cup.
2 The nebulizer is not held an upright position. 3 The item in the cup is too thick to produce aerosol 4 The indoor temperature is too low, fill in 3-6ml hot water (above 80°),DO NOT inhale. |
Low output | 1 Running out of Power, recharge the battery or replace new battery.
2 Check and remove the the bubbles inside the container which are preventing the liquid from coming into continuous contact with mesh plate. 3 Check and remove the residue on the mesh plate, use 2 to 3 drops of white vinegar and 3 to 6 ml of water and run through. DO NOT inhale, rinse and disinfect the container before using again. 4 Mesh plate is worn out and needs to be replaced. |
Which medicines can be used in this nebulizer? | With a viscosity of 3 or lower.For specific liquid for your condition, consult your physician. |
Why is there still liquid in the nebulizer in the end ? | 1 This is normal and occurs for technical reasons.
2 Stop inhaling when the the sound of the nebulizer changes. 3 Stop inhaling when the device automatically shuts due to insufficient inhalant. |
How can this device be used with babies or children? | Cover the baby’s or children’s the mouth and nose with mask to make sure inhale. Note: Children can not allowed to use the device alone, must be done with supervision of an adult. |
Do you need different accessories for different users? | YES, this is very important to maintain proper hygiene. |