Automatic wrist blood pressure monitor U62GH |
Display |
LCD |
Measuring Principle |
Oscillometric method |
Measuring location |
Wrist |
Measurement range |
Pressure:0~299mmHg Pulse:40~199 pulses/min |
Accuracy |
Pressure:±3mmHg Pulse: ±5% of reading |
LCD indication |
Pressure: 3 digits display of mmHg Pulse: 3 digits display Symbol: Memory/Hear beat/Low battery |
Memory function |
2*90 sets memory of measurement values |
Power source |
2pcs AAA alkaline battery DC.3V |
Automatic power off |
in 3 minutes |
Main unit weight |
Appr. 96g( not included batteries) |
Main unit size |
L*W*H=69.5*66.5*60.5mm(2.74*2.62*2.36 inch) |
Battery life |
Could be used for 300 times for normal condition |
Accessories |
Cuff,instruction manual |
Operating environment |
Temperature: 5~40℃ Humidity: 15%~93%RH Air pressure: 86kPa~106kPa |
Storage environment |
Temperature -20℃~55℃,Humidity: 10%~93% avoid crash , sun burn or rain during transportation |
Cuff size |
Wrist circumference appr. size 13.5~21.5cm (5.31~8.46 inch) |
1.Measurement method: oscillometric method
2.Display screen: LCD digital display shows high pressure / low pressure / pulse
3.Blood pressure classification: WHO sphygmomanometer classification indicates blood pressure health
4.Intelligent pressurization: automatic pressurization and decompression, IHB heart rate detection
5.Year/month/day time display
6.2*90sets of measurement results memory for two people; the average reading of the last 3 measurements for data comparison
7.One button measurement,automatic turn on-off for convenient operation
How to set the users?
Press S button when power off, the screen will display user 1/user 2, press M button to switch from user1 to user2 or user2 to user1, then press S button to confirm the user.
How to set the year/month/ date time?
Continue to above step, it will enter into year setting and the screen will flash 20xx. Press M button to adjust the number from 2001 to 2099, then press S button to confirm and enter into the next setting. The other settings are operated like year setting.
How to read memory records?
Please press the M button when power off, the latest 3 times average value will be shown. Press M again to show the latest memory, press S button to show the oldest memory, as well as subsequent measurements can be shown one after the other by pressing the M button and S button each time.